View article: Assessment of the dosimetric impact of intra-fraction motion during frameless treatment delivery on GammaKnife® Icon™.
Assessment of the dosimetric impact of intra-fraction motion during frameless treatment delivery on GammaKnife® Icon™. Open
HDMM tolerances up to and including 3mm all resulted in negligible changes in PCI and GI. Target coverage exhibited greater sensitivity to motion, but only at 3mm was the target coverage reduced below local planning aims. Our HDMM toleranc…
View article: A high-precision, geometric and registration accuracy full-system test method for adaptive SRS, demonstrated on Gamma Knife® Icon™.
A high-precision, geometric and registration accuracy full-system test method for adaptive SRS, demonstrated on Gamma Knife® Icon™. Open
A novel full-system test (FST) phantom and method have been developed to demonstrate and quality assure the geometric accuracy of image co-registration and overall shot delivery in the context of SRS using Gamma Knife® Icon™. The method us…
View article: A standardised method for use of the Leksell GammaPlan Inverse Planning module for metastases.
A standardised method for use of the Leksell GammaPlan Inverse Planning module for metastases. Open
The standardised method for using the optimisation module has potential for shortening treatment times and planning times.
View article: A comparison of the convolution and TMR10 treatment planning algorithms for Gamma Knife<sup>®</sup> radiosurgery.
A comparison of the convolution and TMR10 treatment planning algorithms for Gamma Knife<sup>®</sup> radiosurgery. Open
Convolution provides superior accuracy versus TMR10 in heterogeneous conditions. Implementing convolution would result in increased target doses therefore its implementation may require a revaluation of prescription doses.